Envitec AlcoQuant 6020+
BREATHALYSER
DAILY RATES | ||||
---|---|---|---|---|
Cost per Day | - | - | - | - |
Duration (days) | 1-7 | 8-14 | 15-29 | 30 |
TYPICAL HIRE PERIODS | ||||
---|---|---|---|---|
Cost | $0.00 | $0.00 | $0.00 | $0.00 |
Duration | 7 Days | 14 Days | 29 Days | 30 Days |
Contact us on 1300 601 670 for fantastic long term hire rates.
DESCRIPTION
AlcoQuant® 6020 plus – the reliable companion in all alcohol-screening situations
The new generation with even easier operation and many new functions, With its remarkably intuitive
menu, the new AlcoQuant® 6020 plus is even easier to use. Several new functions and configuration options ensure
that the device is perfectly suited for practical use.
Lightweight, fast, simple, and reliable, The device can be operated with a single hand via three
keys. The entire menu navigation is self-explanatory and immediately understandable with no learning phase. All
messages and readouts are shown on the display in plain language. Optimal readability in both complete darkness and
strong sunlight is guaranteed. The self-monitoring function, ergonomic design, and outstanding sensor technology
ensure reliable results in any situation.
SPECIFICATIONS
Measuring modes: | Active and passive |
Measuring range: | 0 to 5.5 ‰ BAC* |
Accuracy: - Range 0 to 1‰: - Range > 1 to 2‰: - Range > 2‰: |
Max. measuring error in relation
to the ethanol standard (in the
temperature range of -5 °C to 45 °C) ± 0.05 ‰ absolute ± 5 % of measured value ± 10 % of measured value |
Ambient conditions: |
- Temperature range - operation: -10 °C to +50 °C - Temperature range - storage: -25 °C to +70 °C - Optim. storage temperature (sensor): 0 °C to +25 °C - Ambient pressure: 600 to 1400 hPa - Ambient humidity: 20 to 98% r.h. - Minimum expiration volume: 1.2 liters |
Batteries or rechargeable battery (optional): | 4 x LR6 (AA) Alkaline Mignon (recommended: Duracell® Procell®) or EnviteC rechargeable battery NiMH 4.8V (4x2100 mAh) |
Battery charger (optional): |
Protection class II Input: AC 100–240 V/50–60 Hz/93 mA Output: DC 5,6 V/500 mA/2,8 VA |
Charging time: | Approx. 4 hours |
Number of measurements per battery pack/accumulator charging | Approx. 2,000 measurements, depending on battery type and ambient conditions |
Device interface: |
USB and RS-232 interface; data can
be transferred to a PC Optional: Wireless interface (IEEE 802.15.1) for usage with wireless printer |
Dimensions (L x W x H): | 195 mm x 60 mm x 47 mm |
Weight: | 275 g incl. batteries |
Protection class: | IP 54 |
Impact strength: | IK 03 |
Miscellaneous | |
Ready to measure after switching on: | approx. 6 sec. |
Ready to measure after 0‰ measurement: | approx. 2 sec. |
Ready to measure after 1‰ measurement: | approx. 5 sec. |
Sensor type: | E100 electrochemical measuring cell |
Cross sensitivities: | Negligible for other substances, e.g. acetone, etc. |
Measurement units: | Configurable by user: ‰, mg/l (additional upon request) |
Input: | Three-key operation |
Power-saving mode: | Menu settings for automatic device switch-off and backlighting switch-off (“auto-off”) |
Display: | Backlit, fully graphic |
Data storage: | Stores 9,999 measured values with date and time |
Mouthpieces: | Single-use EnviteC mouthpiece (Types S and D) |
Software: | PC software for statistical analysis, data transfer, and logging |
Calibration: | Every six months |
Standard: | DIN EN 15964 |