DESCRIPTION
Fluke 725 Multifunction Process Calibrator “the calibrator” is the one unit that can ensure you are equipped to test and calibrate almost any process parameter. The Fluke 725’s split displays allow you to view both input and output values simultaneously making light work for the user.
Measurement Accuracy
Voltage DC:
30.000 V 0.02% + 2 counts (upper display)
30.000 V 0.02% + 2 counts (lower display)
100.00 mV 0.02% + 2 counts
-10.00 mV to 75.00 mV 0.025% + 1 count (via TC connector)
Current DC:
24.000 mA 0.02% + 2 counts
Resistance 0.0 to 400.0 Ω 0.1 Ω (4-wire), 0.15 Ω (2- and 3-wire)
401 to 1500 Ω 0.5 Ω (4-wire), 1 Ω (2- and 3-wire)
1500 to 3200 Ω 1 Ω (4-wire), 1.5 Ω (2- and 3-wire)
Frequency:
2.0 to 1000.0 CPM 0.05% + 1 count
1.0 to 1100.0 Hz 0.05% + 1 count
1.00 to 10.00 kHz 0.05% + 1 count
Sensitivity 1 V peak-to-peak-minimum
Pressure:
Accuracy from 0.025% of range using any of 29 pressure modules (for detailed specifications refer to pressure modules in options and accessories). Modules available for differential, gauge, vacuum, absolute, dual and high pressure.
Source Accuracy
Voltage DC:
100.00 mV 0.02% + 2 counts
10.000 V 0.02% + 2 counts
-10.00 mV to 75.00 mV 0.025% + 1 count (via TC connector)
Current DC:
24.000 mA (source) 0.02% + 2 counts
24.000 mA (simulate) 0.02% + 2 counts
Resistance:
15.0 to 400.0 Ω 0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA)
401 to 1500 Ω 0.5 Ω (excitation current 0.05 to 0.8 mA)
1500 to 3200 Ω 1 Ω (excitation current 0.05 to 0.4 mA)
Frequency:
2.0 to 1000.0 CPM 0.05%
1.0 to 1100.0 Hz 0.05%
1.00 to 10.00 kHz 0.25%
Waveform 5 V p-p squarewave, -0.1 V offset
Specifications
Ramp functions :
Source functions: Voltage, current, resistance, frequency, temperature
Ramps: Slow ramp, Fast ramp, 25% step-ramp
Loop power function
Voltage 24 V
Accuracy: 10%
Maximum current: 22 mA, short circuit protected
Step functions
Source functions: Voltage, current, resistance, frequency, temperature
Steps: 25% of range, 100% of range